9848FDA 806 Calibration_ Suspend probe at least 1 into a slush of crushed ice and water 32 F for at least 30 seconds Keep the probe from touching the container bottom Press the CAL button for 2 seconds The word CAL appears on display for 2 seconds then displays 32 0 F or 0 C The thermometer is calibrated To avoid accidental re calibration this function can only be performed if the temperature is between 23 F and 41 F 5 C to 5 C Specifications_ Range of temperature measurement Range 40 F to 450 F 40 C to 230 C Accuracy 1 30 F to 180 F 0 5 F 54 to 82 C 0 5 C 2F or 1 C outside of above range Resolution 0 1 F or C up to 199 9 above 1 F and C Auto Shut Off After 1 0 minutes i ...